CALIBRATION
1. Open OTOSUITE
2. Set to Speech
3. Select one wav file & play this file in otosuite. Directly stop the wav file from playing .
4. In the A§E module, calibrate for Generic device, select the correct sound card and transducers in step 1
5. Go to step 2 and play the A§E calibration signal
6. Check in otosuite if the played signal is at 0dB on the input (at our test environment we had to put the slider in A§E to -2.5dB)
7. Set Otosuite as you would play a test and press the run verification button in the calibration settings.
PLAY TEST
1. Open OTOSUITE
2. Set to Speech
3. Set the correct ear
4. Set the correct output (phone, freefield, insert)
5. Set the output to 70dB and start using A§E.
I. if sound does not start in A§E: Play one wav file in otosuite
7. When A§E asks to put the output level to 90dB, please do so in OTOSUITE.
8. After closing A§E, you can also close OTOSUITE.